Metrohm IC Net 2.2 User Manual

Page 98

Advertising
background image

6 Devices

IC Net 2.2

90

The

Manual

tab of the

Metrohm SDU

window is only available for a

connected system.

Pressure

Live display of current pressure in MPa.
This field is also available for the

WATCH

WINDOW

.

Flow

Live display of current flow in mL/min. This
field is also available for the

WATCH WIN-

DOW

.

New flow

A new flow can be set by moving the scroll
bar, by entering a value or by changing the
value using the up and down arrows.
Entry range:

0.01 ... 20.00 mL/min

(depending

on number of pumps)

Min/Max pressure

The

minimum and maximum pressure limit

for the individual pump units can be set in
steps of 0.1 MPa.
Entry range:

0.0 ... 50.0 MPa

.

%B

Individual flow rate for pump B in percent of
total flow.
Entry range:

0 ... 100 %

%C

Individual flow rate for pump C in percent of
total flow.
Entry range:

0 ... 100 %

%D

Individual flow rate for pump D in percent of
total flow.
Entry range:

0 ... 100 %

Advertising