4 hardware settings – Metrohm IC Net 2.2 User Manual

Page 77

Advertising
background image

6.4 761 Compact IC

IC Net 2.2

69

6.4.4 Hardware

settings

Selection of the

Hardware

menu item of the

761 icon

using the right

mouse button opens the hardware settings window consisting of
the three tabs

Hardware

,

Outputs

and

Alarm stops

.

Hardware

The

Hardware

tab of the hardware settings window defines gen-

eral parameters which are set automatically at power on of the
761 Compact IC.

Remote lines after power on

The remote output lines 1

8 are set to this

values after power on or a manual stop with

Shutdown hardware

.

Selection:

0

,

1

Thermostat

Operating temperature of the conductivity
cell.
Selection:

25

,

30

,

35

,

40

,

45 °C

,

off

Thermostatting functions only if the ambient tempera-
ture is at least 5 °C lower than the operating tempera-
ture. It normally takes 30

60 min after power on until

a temperature stability of

±

0.01 °C is attained.

Flow correction

Factor for correction of the difference be-
tween displayed and actual flow rate of the
high-pressure pump.
Range:

0.9

...

1.09

The flow correction is determined by meas-
urement of the actual flow rate with the aid
of a measuring cylinder as follows:

Flow correction

=

Displayed flow rate
Measured flow rate