Metrohm IC Net 2.2 User Manual

Page 86

Advertising
background image

6 Devices

IC Net 2.2

78

Oven

Display of current oven state.

Range

Display of current measuring range.

Full

scale

Display of current full scale range.

Offset

Display of current offset.

Cycle potentials

Graphical display of cycle potentials.

Method

Display of working potentials set in the
Method data frame.

Actual

Display of current working potentials.

<Set>

Send current parameters immediately to the
817 Bioscan. Parameters are not stored in
the system file (

*.smt

) as long as the file is

not saved.


Scan mode

The

Scan mode

subtab of the

Method parameters

tab on the

817

Bioscan

window is used for parameter setting and curve display

for the scan mode. This window is only available if

Mode

has been

set to

Scan

on the

Manual

tab.

Parameters

Method data

Cell potential

Definition of start and end potential for a
user-defined potential sweep, which is ap-
plied to the amperometric detector cell.

E1

Start potential for sweep.
Range:

-2 ... +2 V

.

E2

End potential for sweep.
Range:

-2 ... +2 V

.

Scan cycle

Definition of scan cycle:

full

double

scan

E1 ... E2 ... E1

half

single

scan

E1 ... E2

Selection:

full, half

Scan rate

Definition of scan rate in mV/s.
Selection:

1, 2, 5, 10, 20, 50 mV/s

Oven

Switching on/off column oven and setting
operating temperature.
Range:

15 ... 60 °C

Range

Selection of measuring range:

pA

10 ... 5000 pA

.

nA

0.1 ... 50 nA

.

µA

0.01 ... 5 µA

.