2 general error messages, 3 error messages for 761 compact ic – Metrohm IC Net 2.2 User Manual

Page 42

Advertising
background image

4 Systems

IC Net 2.2

34

Running

Running program or chromatogram data
acquisition.

Running program
(xxx min left)

Running time program with time display.

Waiting for INJECT

Waiting for "INJECT" to start program
and/or chromatogram data acquisition as
defined in the Start mode.

INJECT done

Injection valve has been switched to the
"INJECT" position.

Stopping

Determination has been stopped.

Finished

Program or chromatogram data acquisition
has been finished.

SHUTDOWN

System is shutdown.

4.5.2

General error messages

Detection of hard-
ware failed

Bad connection between PC and device or
device switched off (check connecting ca-
ble or switch on instrument).

4.5.3

Error messages for 761 Compact IC

LEAK DETECTED

The leak detector has detected a leak
(check IC system and connections).

E1

Program checksum wrong (call Metrohm
service).

E2

RAM faulty (call Metrohm service).

E200

Invalid instrument adjustment (call Metrohm
service).

E237

Storage of configuration values failed (re-
peat last action; if error reappears, call
Metrohm service).

E238

Storage of instrument number failed (repeat
last action; if error reappears, call Metrohm
service).

E240

EEPROM faulty (call Metrohm service).

E258

Storage of setup values failed (repeat last
action; if error reappears, call Metrohm ser-
vice).