8 update calibration – Metrohm IC Net 2.2 User Manual

Page 179

Advertising
background image

7.5 Calibration and quantification

IC Net 2.2

171

Base

Base for calculations (

Area

or

Height

) in

quantification and calibration procedures.
Indicates, which parameter is to be used as
a peak response.

Reference channel

Channel which is used to measure area or
height (read-only).

Formula

Formula for calibration function. There are
six possible calibration dependencies for
linear and nonlinear calibration curves.

Statistical weight

Weighting parameter used to weight cali-
bration points deviations on calculating co-
efficients of the calibration formula. Apart
from the uniform weighting

1

the values

1/x

,

1/x

2

,

1/y

and

1/y

2

are also possible.

Standard component

Name of standard component in the case

of

Internal standard calibration

or

Tabulated

calibration

method (details see on-line help).

Concentration of standard

Concentration of standard component in
the case of

Internal standard calibration

or

Tabulated calibration

method (details see on-

line help).

7.5.8 Update

calibration

IC NET / Method / Calibration / Update

Selection of this menu item opens the

Update

window.

Retention time

Replace the expected retention times of the

components with the retention times of the cor-
responding peaks obtained in the current run.

Indexes

Fill

Index

field for components with zero index.

Coefficients

Recalculate all calibration coefficients for all
components (performs recalibration).

Coef. (balance)

Recalculate coefficient of the universal com-

ponent. The aim of this action is to adjust the re-
sponse factor of this component so, that the total