6 Devices

IC Net 2.2

88

automatically switched off. At the same
time, the

Pressure

field color in the 709 set-

tings window and in the

WATCH WINDOW

is

changed to the value set for

Out of range

.

<Set>

Send current parameters immediately to the
709 IC Pump (only available if pump is run-
ning). Parameters are not stored in the sys-
tem file (

*.smt

) as long as the file is not

saved.

<Start>

Set current parameters and start 709 IC
Pump drive.

<Stop>

Stop 709 IC Pump drive.

Links

The

Links

tab of the

709 IC Pump

is used for COM port selection

and settings (details see section 5.2.4 Links).

Advanced

The

Advanced

tab of the

709 IC Pump

window is only available for a

connected system.

Correction factor

Flow correction factor to adjust the dis-
played flow to the actual flow measured
with the aid of a measuring cylinder.
Entry range:

0.90 ... 1.10

<Ask>

Read

current

Correction factor

value.

<Set>

Send

current

Correction factor

value to the

709 IC Pump. This parameter is not stored
in the system file (

*.smt

) as long as the file is

not saved.