7.6 Report output

IC Net 2.2

177

4 11.19 5.0 9.933e-03 0.00 0.000 0 bromide

5 14.60 5.0 7.912e-03 7.56 0.000 S 0 nitrate
6 16.80 5.0 1.556e-02 0.00 0.000 0 phosphate
7 21.26 5.0 6.021e-03 5.57 0.000 0 sulphate

Calibration results

Calibration results of the component win-
dow. For each component a new page with
calibration results and calibration curve is
put out.

CALIBRATION OF COMPONENT fluoride

Method: Multicalibration.mtw
Equation: Q = - 7.48163·A^2 + 0.00262326·A + 0.0233088
RSD: 0.691 %
Correlation coefficient: 0.999277


K3 = 0 K2 = -7.48163e-08 K1 = 0.00262326 K0 = 0.0233088
Base: Area
Ref.channel: ch1
ISTD:
Formula: Y=K2·X^2+K1·X+K0
Weight: 1

Level Height Area Conc. Volume Retention Used File

1 6.46 173 2 1 10.8 Yes g8140800.chw
2 34 897 10 1 10.8 Yes g8140823.chw
3 90.9 2.36e+03 25 1 10.8 Yes g8140858.chw
4 198 5.04e+03 50 1 10.8 Yes g8140929.chw

Channel table

Parameters of the data acquisition channel
used.

CHANNELS TABLE
ch1
No Name Units Input Minimum Zero Maximum Range Coefficient Noise Shift

1 ch1 mV 1 -8388607 0 8388607 2.500e+03 2.980e-04 4.15 0

# Name/Units Noise RMS PeakToPeak Drift/hour
1 ch1 4.2 29976.5 330509.6 3585.801
mV 0.00124 8.93 98.5 1.07

Name

Channel’s name. This text appears as

Channel label

in the chromatogram.

Units

User-defined units of detector response.

Input

Number of data acquisition channel.

Minimum

Minimum value of the linear range of the AD
converter (in ADC conversion units). This