Metrohm IC Net 2.2 User Manual

Page 160

Advertising
background image

7 Methods

IC Net 2.2

152

European Pharmacopoeia

Formula setting according to European
Pharmacopoeia.

US Pharmacopoeia

Formula setting according to US Pharma-

copoeia.

Void time/volume

Calculation of void time/void volume:

Calculation method

Selection of the method for void time calcu-

lation:

None

Manual entry in the

Void time

field.

1st

component

The peak corresponding to the first com-

ponent is selected as a void time marker
and its retention time replaces the previous
value of the void time. If the first component
is not identified, the expected retention time
is used.

1st peak

The first detected peak is used as a void
time marker for the run and its retention
time is stored in the

Void time

field.

From void volume %

For the calculation of the void time, the %
value entered in the

Void volume

field is mul-

tiplied by the ratio of empty column volume
and eluent flow rate.

Void volume

Void volume in mL and % of column vol-
ume. The % value can be entered manually
if

Calculation method =

From void volume %

is

set.

Void time

Void time in s. This value can be entered
manually if

Calculation method =

None

is set.

Index

Calculation of Retention index:

Interpolation

Use of linear or logarithmic interpolation
scale for retention index calculation.

Internal

The index scale is constructed on the basis
of the current chromatogram. All compo-
nents used for index scale calibration
should be present in the current sample.

External

The index scale is constructed on the basis
of another standard chromatogram.