ITC IRMA TruPoint Blood Analysis System User Manual

Page 100

Advertising
background image

7.4

5. Enter the Cal code and press

next

to return to

the Lot picklist. Press

done

to return to TEST

SETTINGS.

Note:

• New product lots that are entered via patient test or QC test will appear

in the Lot picklist following entry.

Replacing an existing product lot with a new product lot

1. Highlight

Lot Entry

and press

edit

. Highlight the appropriate product type

and press

next

. Scroll down the picklist until the existing lot number is

highlighted. Press

edit

.

2. Enter the new lot code and press

next

. Enter the new Cal Code and press

next

to return to the Lot picklist. The new lot replaces the old lot in the

picklist. Press

done

to return to TEST SETTINGS.

CALCULATIONS

QA Users can define the calculated values to be reported with each test, and
the formulas used to derive the calculated values using the Calculations
option.

Defining the calculated values to be reported with each patient test

1. Highlight

Calculations

and press

edit

.

2. Highlight

Calculated Values

from the picklist

and press

edit

.

3. Highlight the calculated values to be reported

(depends on product type) with each patient
test, and press

next

.

If BEb was not selected, Calculated Values
setup is complete.