7.17

Setting up or modifying QC lockout schedule

4. The

Analyte Lockout screen displays the

current settings for the first analyte/analyte
group available on the product type selected.
Press “on” to activate or “off” to deactivate
QC Lockout for the selected analyte/analyte
group. If “on” selected:

a. Press

next

to keep current settings and advance to the QC lockout

settings screen for the next analyte available on the product type.

b. Press

edit

to change settings.

5. Select shift length (8, 12, or 24 hours) from the

Shift Length screen. Press

next

.

6. Select Shift 1 Start time using the numeric and

colon keys. Press

next

.

7. The screens that follow will depend upon

whether EQC or Liquid QC Lockout is being
configured:

a. EQC Lockout: Select Shifts for Lockout

(1, 2, or 3) by highlighting the

yes

button

next to desired shifts. Press

next

to return to

the Select Product Type screen.

b. Auto EQC Printout: The auto EQC report

can be printed in a detailed (full) or
abbreviated (short) report. The short report
provides a printout indicating pass or failed
EQC. For a detailed report, press “Full” and
press “Short” for an abbreviated report.

c. Liquid QC Lockout: Select the number of

controls to be run on shift 1 (1, 2, or 3).
Press

next

. Repeat step to define number of

controls to be run on each shift. Press

next

to

advance to lockout setup for each analyte.

CONTROLS

QA Users can set up liquid controls for each product type. The analyzer will
store a maximum of 42 individual control entries for cartridge product types.
Once this limit is reached, an existing control must be deleted before a new
control can be added.