ITC IRMA TruPoint Blood Analysis System User Manual

Page 29

Advertising
background image

1.21

3. a. If the Communications Method setting is

User Selects, highlight the button next to
the appropriate method and press

next

and

go to step 4.

b. If the Communications Method setting is

not User Selects, go to step 4.

4. Connect cable to computer and verify that

DeviceCom is running. Select an option on
the Device Update screen:

a. “receive all” - All profile settings will be sent to the analyzer,

overwriting the existing settings. If new software or language updates
are available, they will be also be sent.

b. “receive new” - Only profile settings that have changed since the last

analyzer update will be sent to the analyzer. If new software or
language updates are available, they will be also be sent.

5. The available updates (i.e., Languages,

Software, Settings, None), and the estimated
time it will take to complete the update will
display. If there is a profile associated with
the analyzer and there are

no unsent records in the analyzer, pressing

ok

will continue the update

until completion.

If the update cannot proceed for one of the following reasons, a message
screen will be displayed:

a. The analyzer is not associated with a profile.

> Press

ok

to continue update. Factory

default IRMA TruPoint settings will
overwrite existing analyzer settings. If
new software or language updates are
available, they will be also be sent.

> Press

cancel

to cancel update and return to the SETTING OPTIONS

menu.

b. Result Transfer setting is “idms” and

there are unsent results.

> Unsent results will automatically be

transferred when

ok

is pressed from the

Updates Available screen.