ITC IRMA TruPoint Blood Analysis System User Manual

Page 27

Advertising
background image

1.19

Transmitting Results to VueLink

1. The patient test results displayed on the

Results screen are transmitted to VueLink
when

done

is pressed, only under the

following conditions:

• The VueLink option in DEVICE SETTINGS is “on”, and;

• The IRMA TruPoint analyzer is connected to a VueLink module

residing in a patient monitor or monitor rack. If there is no connection
between IRMA TruPoint and VueLink, pressing

done

will end the test

and advance IRMA TruPoint to the appropriate screen.

2. The

Sending to VueLink screen displays once

a connection has been established. The user
has 15 seconds to verify that the patient ID
from the test record transmitted matches the
patient ID entered into the patient monitor.

a. Patient ID verified within allotted time - test results display on the

patient monitor and the IRMA TruPoint advances to the appropriate
screen, depending on setup.

b. Patient ID not verified within allotted time - test results do not

display on the patient monitor and the IRMA TruPoint advances to the
appropriate screen, depending on setup.

Transmission Errors

If a connection is made, but the transmission is
unsuccessful, an error message will display.

a. Resend - verify connections and receiving

system and press

resend

button.

b. Cancel transmission - press

cancel

button.

Transmitting Recalled Results

All IRMA TruPoint patient test results can be recalled and transmitted to
VueLink.

• A result can be transmitted multiple times.

• A result cannot be edited in IRMA TruPoint once it has been transmitted

to VueLink.

• Transmitting a result to VueLink does not affect its sent/unsent status for

transmission to idms or a host system.

Advertising