7.18

QC limits may be entered into the analyzer for each control (optional). These
limits are used during quality control testing to verify that the analysis system
is performing properly. QC results will be flagged only when limits are
defined.

Setting up a new control

1. Press

QC

on the SETTING OPTIONS menu.

2. Press

Controls

on the QC SETTINGS menu.

3. Highlight the product type of choice. Press

next

.

4. The

Controls screen displays a picklist of current controls. Press

add

for a

new control.

5. Select Level, (1, 2, or 3). Press

next

.

a. If the bar code scanner is attached to IRMA

TruPoint, the Please scan control screen
will appear.

>

Scan the barcodes from the expected
values sheet. These sheets are available
at www.itcmed.com.

>

The touchscreen will track bar code
scan progress.

b. If the barcode scanner is not being utilized, proceed as follows:

6. Enter Control Lot (1-7 alpha-numeric characters) for the new control.

Press

next

.

7. Enter Expiration Date (MM/YY). Press

next

.

8. Select Analytes to be reported for the new control by highlighting all

desired analytes. Press

next

.

9. Follow the series of Limits entry screens to establish control limits for

each of the selected analytes. The product type, control level, and control
lot number appear on each screen.

10. Enter numeric lower and upper limits, or press

next

through each screen

to enter no limits.