ITC IRMA TruPoint Blood Analysis System User Manual

Page 41

Advertising
background image

2.11

Note:

• If air bubbles or calibration gel are present in the sample path after the

initial sample injection, remove them by injecting additional sample from
the same syringe, or by lightly tapping the top of the syringe plunger and
then injecting additional sample from the same syringe.

• If air bubbles or calibration gel cannot be removed from the sample path,

press

cancel

to stop the test, discard the cartridge, and begin again with a

new cartridge.

Proper injection technique will prevent air bubbles or calibration

gel from being present in the sample path after initial sample injection.

20. Ensure that no air bubbles or calibration gel are present in the sample

path, then immediately press

test

to continue sample analysis.

a. The Analyzing screen displays if there is no

required patient bypass status or patient ID
information to be entered.

b. If patient bypass status or patient ID

information entry is required, but not yet
entered, the Bypass Status screen or
Patient ID screen will display when

test

is

pressed. The required information must be
entered

before results will display or print. Refer to

Section 2-Entering Test Information and
Section 2-Patient Bypass Status for details.
Enter the required information and press

next

to advance to the Analyzing (or

Results) screen.

21. When the test is complete, remove the cartridge with the collection

device attached. Dispose of both in accordance with established
guidelines for your facility.

View Test Results

22. Results automatically appear on the IRMA TruPoint touchscreen when

the analysis is complete. Press

more

to view additional measured or

calculated results if all the results do not appear on the screen.