ITC IRMA TruPoint Blood Analysis System User Manual

Page 24

Advertising
background image

1.16

Bar Code Scan Symbologies

The bar code reader is factory programmed to read the following bar code
symbologies:

Code 11

EAN/JAN

Plessey

Code 39

Interleaved 2 of 5

Telepen

Code 128

Matrix 2 of 5

UPC

Codabar MSI

Scanner Specifications

Refer to the User’s Guide that came with the bar code reader for detailed
specifications and regulatory compliance statements.

Note:

• The Welch Allyn bar code reader uses a non-laser, red LED light source.

Connecting the Bar Code Reader

1. Ensure that the IRMA TruPoint analyzer is turned “off”.

2. Connect the bar code reader cable to the round keyboard port on the back

of the analyzer.

3. The reader will be ready for use when the analyzer is powered back “on”.

Caution:

• The IRMA TruPoint analyzer can only determine the presence or

absence of a bar code reader during analyzer power-up. Do not connect
or disconnect the reader when the analyzer is “on”. Doing so may
result in damage to the reader or analyzer, or the display of incorrect
screen prompts.

Disconnecting the Bar Code Reader

1. Ensure that the IRMA TruPoint analyzer is turned “off”.

2. Disconnect the bar code reader cable from the keyboard port on the back

of the analyzer.

Mounting the Bar Code Reader

The bar code reader may be attached to the IRMA TruPoint analyzer using
the mounting clip that came with the reader.

1. Attach the clip to the IRMA

TruPoint handle as shown.
(Figure 1-6).

Figure 1-6

Advertising