C.6

Patient Temperature Correction

7, 8, 9, 10

Patient temperature can be entered into the IRMA TruPoint analyzer if it
deviates from 37°C. The measured pH and blood gas values are recalculated
at the input temperature (T) with the following equations:

pH

(T)

= pH

(37°C)

- 0.0147(T-37) + 0.0065(7.40 - pH

(37°C)

)(T-37)

pCO

2(T)

= pCO

2(37°C)

x 10

(0.019{T-37})

Severinghaus pO

2

equation:

5.49*10

-11

*pO

2

3.88

+0.071

9.72*10

-9

*pO

2

3.88

+2.30

( Temp)

pO

2Final

= pO

2

* 10

Temp is T-37°

Kelman-Nunn pO

2

equation:

pO

2Final

= pO

2

x 10

[{0.0052 + 0.027(1-10^-0.13(100-O2Sat))} ( Temp)]

Temp is T-37°

MDRD GFR

MDRD GFR (mL/min/1.73m

2

)=186 X (CR

*

)

-1.154

x (age

**

)

-0.203

X (0.742 if

female) x (1.21 if black) mg/dL

*

= mg/dL

**

= years

Advertising