Test settings – ITC IRMA TruPoint Blood Analysis System User Manual

Page 98

Advertising
background image

7.2

TEST SETTINGS

QA Users can access Test Settings by pressing

Test

from the SETTING OPTIONS menu. A TEST

SETTINGS picklist displays. Current test settings
can be viewed or changed by highlighting a
category in the TEST SETTINGS picklist and
pressing

edit

.

PRODUCT SETUP

QA Users can use the Product Setup option to:

Define the product types that are in use at their facility.

With each

patient test, the user is prompted to enter the product type being run. Only
those product types defined in the Product Setup option will appear on the
Select Product Type screen with each test; if a single product type is
defined, the analyzer assumes that all cartridges inserted for testing are of
that type, and no Select Product Type screen will appear.

Define the analytes that will be reported for each cartridge type.

Note

that all analytes will always be reported for BG cartridges.

Defining the product type(s) in use

1. Highlight

Product Setup

in the picklist and

press

edit

.

2. Highlight each Product Type that will be used

at your site. Press

next

to define which

analytes will be reported for each cartridge
type.

Defining analytes to be reported for each cartridge type.

Define which analytes will be reported for each CC test:

1. Highlight the analytes to be reported

automatically with each CC test. Press

next

.

2. If the analytes to be reported will vary from

test-to-test, highlight

User Selects

. This will

prompt individual selection of analytes or
analyte groupings with each CC test. Press

next

.