ITC IRMA TruPoint Blood Analysis System User Manual

Page 47

Advertising
background image

2.17

Patient Status

• Highlight the appropriate Patient Status setting.

› If Code or Room Air status is selected,

press

next

to accept the selected oxygen

status and complete entry.

› If Pump status is selected, press

next

to access the FIO

2

screen. Enter

FIO

2

value and press

next

to access Comments screen. Enter a one- to

twelve-character comment (optional) and press

next

to complete entry.

› If the Oxygen status is selected, press

next

to access the Oxygen

Therapy screen.

Oxygen Therapy

• Highlight the appropriate Oxygen Therapy

mode. Press

next

.

• The entry options displayed depend on the

relevant settings for the selected oxygen
therapy mode.

Ventilator Mode

• The

Ventilator Mode screen appears following

selection of Vent. from the Oxygen Therapy
screen. Highlight the appropriate mode and
press

next

.

• The following information may be entered, depending upon the ventilation

mode selected:

Tidal Volume (mL)
Rate or Frequency (B/min.)
FIO

2

(%)

PEEP (cm H

2

O)

IPAP (cm H

2

O)

EPAP (cm H

2

O)

Comments (up to twelve alpha-numeric characters)

Mask Type

• The Mask Type screen appears following

selection of Mask from the Oxygen Therapy
screen. Highlight the appropriate mask type and
press

next

.