Temperature test – ITC IRMA TruPoint Blood Analysis System User Manual

Page 65

Advertising
background image

3.13

TEMPERATURE TEST

PERFORMING A TEMPERATURE TEST

Turn IRMA TruPoint Analyzer On

1. Touch the right-hand edge of the screen to turn the analyzer “on”.

Do NOT insert a cartridge to power-on the analyzer.

Note:

• All QC tests must be initiated via the main MENU, QC Test option. If a

cartridge is inserted to power-on the analyzer, or inserted at the main
MENU screen, the analyzer will automatically initiate a patient test, not
a QC test.

Enter User ID (optional)

2. If the User ID option has not been activated, go to step 4.

3. If the User ID option is activated, the Enter or Scan ID screen will

display. The User ID may be entered using either the alpha/numeric
keypads or the bar code scanner.

a. Entry via touchscreen keypads - enter your User ID and press

next

.

b. Entry via bar code scanner - scan your User ID. The IRMA TruPoint

will automatically advance to the next screen.

Initiate Temperature test

4. Press

QC Test

on the main MENU. The QC

TEST OPTIONS menu is displayed.

5. Press

the

Thermometer

on the QC TEST

OPTIONS menu to initiate a temperature test.

Insert Temperature Card

6. Fully insert the Temperature Card into the

analyzer.

Verify Card Information

7. Verify that the Temperature Card Serial

Number and Cal Code match the information
displayed on the Verify Information screen.

Advertising