ITC IRMA TruPoint Blood Analysis System User Manual

Page 73

Advertising
background image

4.7

b2. The Select Control screen displays a

picklist with the control lots and levels
available for the selected Product
Type. The analytes available and
control expiration date are

shown below the picklist for the highlighted control. Highlight the
desired control and press

next

.

b3. Enter first and last search dates.

b4. The Output Type screen allows

selection of

Result

(detailed report) or

Table

(summary report).

“Result” Output Selected:

The

QC Results Found screen displays a

picklist with test dates and times. Scroll
through the list of results until the
desired result is highlighted. The
product type and QC lot, level, and
analytes are shown below the picklist

for the highlighted result. Press

print all

to print a detailed hard copy for all
results found in the picklist. Press

view

to

see the results for the highlighted result
only. The options from the QC Results
screen are:

print:

Prints a hard copy of the QC test results selected.

limits:

Displays QC limits, if they have been established

by the QA User.

done:

Returns to QC RESULTS FOUND menu.

“Table” Output Selected:

The

Verify Results to be Printed screen

displays. The options from this screen
are:

print:

Sorts results and prints a table

for all the results that meet the search criteria displayed. Printing

can be stopped by pressing the

stop

button, and restarted by

pressing

resume

. Upon successful completion of printing, the

system returns to the SEARCH QC RESULTS menu.

edit

: Returns to SEARCH QC RESULTS menu.