ITC IRMA TruPoint Blood Analysis System User Manual

Page 32

Advertising
background image

2.2

CAPILLARY REQUIREMENTS

Capillary samples must be collected in the IRMA TruPoint Capillary
Collection Device. See Appendix F for detailed Capillary Collection Device
instructions.

SAMPLE SIZE

The minimum whole blood sample volumes that must be injected into a
single-use cartridge for a patient sample analysis are:

200 uL

if sample collected in a syringe

125 uL

if sample collected in an IRMA TruPoint Capillary Collection

Device.

• Ensure that sufficient sample is collected to meet the minimum volumes

required for injection into the cartridge.

GENERAL SAMPLE COLLECTION GUIDELINES

• Time sample collection to minimize delays between collection and

analysis.

• Avoid drawing samples above an IV line to prevent dilution of the sample

with IV fluid.

• When drawing a sample from an indwelling line, back-flush and clear the

line of IV fluids prior to sampling to remove anticoagulants or medications
which might interfere with the test.

• Allow the blood collection site to dry after being cleansed with alcohol to

prevent hemolysis.

• Fill the collection device to the appropriate capacity. Incomplete filling

may cause high heparin to blood ratios which may lower ionized calcium
results and may affect other results.

• Thoroughly mix samples collected in syringes.

BLOOD GAS SAMPLE HANDLING

• Expel any air present in the syringe immediately after collection and

before the sample is mixed. If a portion of the sample must be separated
for other testing, do not expose sample to air.

• If a sample cannot be tested within 5 minutes of collection:

> Expel all air from the syringe.

> Cap or seal-off the end of the collection device.

> Store the blood gas sample in an ice slurry.

> Thoroughly mix the sample while the cartridge is calibrating.