7.8

Setting up Patient Notes:

1. Turn Patient Notes “on” and press

next

.

2. Default or established patient notes appear in a picklist.

a. Deleting Patient Notes: Highlight the

desired note and press

delete

.

b. Adding Patient Notes: Press

add

and enter

a new note via the touchscreen keypad.

Patient notes may be up to 12 characters in length, and a maximum of

15 patient notes may be established.

Setting up QC Notes:

1. Turn QC Notes “on” and press next.

2. Default or established QC notes appear in a picklist.

a. Deleting QC Notes: Highlight the desired note and press

delete

.

b. Adding QC Notes: Press

add

and enter a new note via the touchscreen

keypad. QC notes may be up to 12 characters in length, and a
maximum of 15 QC notes may be established.

CORRELATION

QA Users can establish correlation parameters for each IRMA TruPoint
product type to adjust IRMA TruPoint patient test results to values that would
be expected if samples were run on another analyzer (i.e., designated
reference analyzer). These correlation factors are based on split-sample blood
results from a statistically valid number of samples. The IRMA TruPoint
analyzer can be run without using the Correlation feature, even if correlation
values have been entered into the analyzer.

There are 2 correlation options:

Off:

Correlation parameters will not be used by the analyzer to adjust

IRMA TruPoint patient test results.

On:

Correlation parameters will be used by the analyzer to adjust IRMA

TruPoint patient test results.