Patient test procedure – ITC IRMA TruPoint Blood Analysis System User Manual

Page 36

Advertising
background image

2.6

5. Once the sample path is completely filled with a minimum of 200 uL of

sample, press

test

to begin sample analysis. Leave the syringe attached to

the cartridge until the analysis is complete.

INJECTING A CAPILLARY SAMPLE

The IRMA TruPoint Capillary Collection Device must be used to collect and
inject samples into IRMA TruPoint cartridges. Refer to Appendix F of this
manual, or the IRMA TruPoint Capillary Collection Device package insert for
instructions.

PATIENT TEST PROCEDURE

PERFORMING A PATIENT TEST

Initiate a Patient Test

There are two ways to begin a patient test:

1. Touch the right-hand edge of the screen to turn the analyzer “on”; or

2. With the analyzer “off”, insert a cartridge to turn the analyzer “on” and

automatically initiate a patient test (see steps 7-10 for cartridge insertion
instructions).

Enter User ID (optional)

3. If the User ID option has not been activated, go to step 5.

4. If the User ID option is activated, the Enter or

Scan ID screen will display. The User ID may
be entered using either the alpha/numeric
keypads or the bar code scanner.

a. Entry via touchscreen keypads - enter your User ID and press

next

.

b. Entry via bar code scanner - scan your User ID. The IRMA

TruPoint will automatically advance to the next screen.

5. a. If the analyzer was powered “on” by

touching the screen, the main menu will
display. Press

patient test

.

a1. If no bar code reader connected the

Select Product Type screen displays.
Go to step 6

a2. If bar code reader connected - the Scan cartridge bar code screen

displays. Scan the cartridge package label for the cartridge to be
used. The cartridge product type, Lot Code, and Cal Code are
encoded in the single bar code on the cartridge label. The IRMA
TruPoint will automatically advance to the next screen (the Select
Product Type
and Verify Information screens will not be