ITC IRMA TruPoint Blood Analysis System User Manual

Page 77

Advertising
background image

4.11

2. The

Automatic Results Transfer option is

configured “on”.

3. The IRMA TruPoint analyzer is operated with

the AC adapter. Automatic Results Transfer
will not initiate when the analyzer is powered
by battery.

4. The IRMA TruPoint analyzer is connected to the receiving system via

the appropriate cable.

After a four to six minute period of inactivity on the analyzer, the IRMA
TruPoint will make a single attempt to transmit results. If all of the above
criteria have been met, the IRMA TruPoint will begin to transfer results and
the Results Transfer in Progress screen will display. Data transfer can be
stopped while in progress by pressing

stop

; the screen will return to the main

MENU or Enter ID screen.

Note:

• See

the

idms User Manual for idms setup instructions and cable and

adapter information.

• See the DeviceCom User Manual for DeviceCom setup instructions.

• See the External Modem User Manual for setup instructions.