5.7

SCREEN MESSAGE

REASON FOR

MESSAGE

CORRECTIVE ACTION

A

NALYZER MEMORY

“Attention! Approaching

result capacity”

Patient, EQC, LQC, or
Temperature Test was
selected and the analyzer
memory was within 20
records of being full.
Message will display only if
results transfer is “on”.

If Auto-EQC is “on” the
message will display when
the analyzer exits sleep
mode.

Press ok to continue test. Following
completion of the test, transfer stored
data if stored results should not be
overwritten. See Section 4-Data
Access
. Once storage capacity is
reached, analyzer will begin to
overwrite stored results.

A. Press cancel to abort test if stored
results should not be overwritten.
Transfer data before continuing.

“WARNING! Reached

result capacity”

Patient, EQC, LQC, or
Temperature Test was
selected and the analyzer
memory was full. Stored
results will be overwritten.
Message will display only if
results transfer is “on”.

If Auto-EQC is “on” the
message will display when
the analyzer exits sleep
mode.

B. Press ok to continue the test (results
that have not been transferred may be
overwritten).