7.14

REPORTABLE RANGES

The default reportable range for each analyte can be found in Appendix B,
Table B-8. Due to regulatory requirements, some facilities may want to
specify more narrow reportable ranges for certain analytes. QA Users can
establish a custom reportable range consisting of an upper and lower limit for
each analyte.

Note:

• Changing reportable ranges may affect stored records. Stored results

should be uploaded to idms before reportable ranges are changed.

• An error message will appear upon entry of a reportable range that

conflicts with established QC limits or reference ranges (e.g., a QC or
reference range limits cannot be outside a reportable range limit).

Changing reportable ranges

1. If applicable, upload stored results to idms and confirm that existing QC

limits and reference ranges do not conflict with the reportable ranges that
will be entered.

2. Highlight

Reportable Ranges

from the TEST SETTINGS menu. Press

edit

.

3. Highlight the appropriate button:

a. Select

Default

to set analyzer to default reportable ranges for all

analytes. Press

next

to complete entry.

b. Select

User Defined

to enter custom reportable ranges. Press

next

. An

analyte picklist is displayed.

4. Select the desired analyte from the picklist and press

edit

. If reportable

ranges will be changed for multiple analytes, select “All”.

5. Enter the lower reportable range limit and press

next

.

6. Enter the upper reportable range limit and press

next

.

7. Continue for all appropriate analytes.