ITC IRMA TruPoint Blood Analysis System User Manual

Page 66

Advertising
background image

3.14

a. If the information displayed matches the information on the

Temperature Card, press

next

. Proceed to step 8.

b. If the information displayed does not match the information on the

Temperature Card, press

edit

to enter or select the correct serial

number and Cal Code. Verify the information and press

next

.

8. The

Analyzing screen appears, indicating that the temperature test is in

progress.

Review Results

9. Once the temperature test is complete, the analyzer displays the

Temperature Test Results status screen. The result for a temperature test
is either “Passed” or “Failed”.

a. Passed Temperature Test:

a1. Press

notes

to select QC notes to attach

to the record (optional).

a2. Press

print

to obtain a printout.

(Results may automatically print,
depending on the auto-print setup).

a3. Press

done

to return to the QC TEST OPTIONS screen.

b. Failed Temperature Test:

b1. Press

help

for assistance in resolving

the problem that caused the failure
(optional).

b2. Press

notes

to select QC notes to attach

to the record (optional).

b3. Press

print

to obtain a printout. (Results may automatically print,

depending on the auto-print setup).

b4. Press

done

to return to the QC TEST OPTIONS screen.

REFERENCES

1. Elin RJ: Elements of cost management for quality assurance. College of

American Pathologists, pp 182-183, April 1980.

Advertising