4.1

Section 4

Data Access

OVERVIEW

This section describes the IRMA TruPoint data storage capabilities, the
procedures for patient and QC test result recall, and result transfer to idms or
other connected system.

DATA STORAGE

Results can be recalled, printed, and/or transferred to other systems. The
analyzer stores the following number of test records even when powered-off
or when the battery is removed:

200 patient test records

150 liquid QC records

100 EQC/Temperature Test results

• Error codes for the last 10 failed EQC/Temperature Tests

Access the RECALL menu by pressing the

Recall

button on the main MENU.

The following options are presented:

Patient:

Search for patient test records by patient ID and/or date range, or
recall results for the last test performed. Results and associated test
information can be viewed and/or printed. Test information can be
modified when recalled by the last result performed option if the
record has not been sent to idms, VueLink, or an LIS/HIS.

QC:

Search for QC test records by QC type, product type, control
lot/level
, and/or date range, or recall results for the last QC test
performed. QC Notes can be modified when recalled by the last
result performed option.

Send:

When the results transfer option is “on”, new or previously

transferred results can be sent via the RS232 port, LAN 10/100, or
modem to another system capable of accepting the ASTM output
(i.e., idms or other information systems capable of accepting ASTM
formatted data).

Status:

View the battery information and About IRMA TruPoint screens.
This information may be requested by your service provider to help
resolve operating problems.