B.1

Appendix B

System Specifications
and Cartridge
Information

This appendix includes IRMA TruPoint system specifications and IRMA
TruPoint cartridge information.

SYSTEM SPECIFICATIONS

General Specifications

Table B-1

Category Specification

Operating Temperature:

12–30°C (54-86°F)

Measurement Temperature:

37°C (Blood gases)

Analyzer Shipping/Storage
Temperature:

0-50°C

Operating Relative Humidity:

0–80% (non-condensing)

Operating Barometric
Pressure:

350–900 mmHg (measured by an
onboard barometer)

Sample Size:

0.125 –5.0 mL

Analysis time:

Approximately 60 to 300 seconds after
sample injection (depends on
parameters)

Size (LxWxH) and Weight:

Cartridge/Temperature Card:

2.6” x 1.3” x 0.6”; 0.5-0.7 oz.

66 x 33 x 15.24mm; 15-21 g

Analyzer:

11.5” x 9.5” x 5”; 5 lbs., 4 oz.

292.1 x 241.3 x 127mm; 2381g

Battery Charger/Power Supply:

7.5” x 3.5” x 3.5”; 2 lbs.

190.5 x 89 x 89mm; 907g

Rechargeable Battery:

6.25” x 2.25” x 1.25”; 14 oz.

159 x 57 x 32mm; 397g