6 digital i/o – Rice Lake iQUBE2 Digital Diagnostic Junction Box User Manual

Page 15

Advertising
background image

Installation

11

2.6

Digital I/O

Each iQUBE

2

board provides four channels of digital I/O.

Digital inputs can be set to provide many indicator functions, including four keys (

Zero

,

Tare

,

Units

,

Mode

) plus cell

health, and setpoints (see Table 5-10 on page 45 for a complete listing). The four digital inputs/outputs on the pri-
mary CPU are configurable; any secondary CPU board’s I/O will be used for cell health only.

The fastest I/O reaction times are achieved when the iQUBE

2

controls its own free-running setpoints using its own

local I/O (2 milliseconds). If preacts are required, the 920i should be used as the controller for its own I/O.

RLWS does not recommend the use of iQUBE

2

(local) I/O with 920i controlled setpoints using the preact functions

as variances in the reaction time in serially controlled I/O can dramatically affect the preact algorithm.

Cell Health

The iQUBE

2

CPU board provides bi-color cell status LEDs next to each load cell connector. These LEDs indicate

the status of connected load cells.

Digital inputs are active low (0 VDC), inactive high (5 VDC). Digital I/O are active/low if the CPU board’s LED
color is green.
Digital outputs are typically used to control relays that drive other equipment. Outputs are designed to sink, rather
than source, switching current.
Each output is a normally open collector circuit, capable of sinking 25 mA when active. Digital outputs are wired
to switch relays when the digital output is active (low, 0 VDC) with reference to a 5 VDC supply.
Table 2-10 shows the pin assignments for connector J5.

Data Rate

Approx. Range

54 Mbps

65 feet

48 Mbps

80 feet

36 Mbps

115 feet

24 Mbps

140 feet

18 Mbps

170 feet

11 Mbps

180 feet

9 Mbps

190 feet

5.5 Mbps

215 feet

3 Mbps

260 feet

1 Mbps

330 feet

Table 2-8. Approximate 802.11g data rate/range

LED Color

Meaning

Green

Load cell is good

Red

Load cell error

Red/Green

Flashing

Consistent error being cleared by host

Off

Disabled/load cell not assigned to a scale

Table 2-9. LED Status Indications