3 section match, Section match – Rice Lake iQUBE2 Digital Diagnostic Junction Box User Manual

Page 42

Advertising
background image

38

iQUBE

2

Installation Manual

4. Select whether you want standard calibration with or without linear points, then click

Next

.

5. Enter the amount of test weight being used.
6. If certified test weights are being used during cell normalization, select the check box. This will calibrate the

scale during normalization.
If chains or hooks are being used during the calibration, select the check box.

7. Remove all weight from the platform and click

Calibrate Zero

. If you want to use the zero acquired from the

last press of the Zero button, click

Use Last Zero

.

The status bar (lower left corner) of the window displays Transmitting... until calibration is finished. It then
displays Zero Calibration Complete.

8. Click

Next

.

The Corner Match Calibration Wizard displays all load cells installed on the system.

Figure 4-5. Corner Match Calibration wizard

9. Place the test weight whose value you entered in step 5, over the specific load cell.
10. Click

Measure

for the load cell corresponding to the corner with weight on it.

Under Status,

Success

will display when measurement is finished. As each measurement is completed, the

next cell will become available to measure.

11. Repeat steps 9-10 until each cell has been measured.
12. Click

Next

.

13. If certified test weights are not being used, a span calibration must now be completed. If they were being used

(see Step 6), calibration was completed during normalization.

14. Add the test weight to the scale and click

Calibrate Span

.

15. If linear points were selected in Step 4, enter the applied weight in the text box and click

Measure

.

The Status will display

Success

when complete.

16. Click

Next.

17. If chains or hooks are being used, remove them (along with all other weight) from the scale and click

Re-Zero

Calibrate

. Click

Next

.

18. Calibration is complete. Click

Finish

to commit the calibration values and send them to iQUBE

2

.

Click

Cancel

if you want to abort. This will require a re-calibration at a later time.

4.2.3 Section Match

This option measures and trims load cell values in platform section pairs, and allows cell trimming.

1. From the

Scales

section, select the scale to be calibrated.

2. From the

General

tab, select

Section Match

from the

Calibration Type

pull-down menu.

3. Click

Calibrate

.

The Section Calibration wizard appears.

4. Select whether you want standard calibration with or without linear points, then click

Next

.

5. Enter the amount of test weight being used.
6. If certified test weights are being used during cell normalization, select the check box. This will calibrate the

scale during normalization.
If chains or hooks are being used during the calibration, select the check box.

7. Remove all weight from the platform and click

Calibrate Zero

. If you want to use the zero acquired from the

last press of the Zero button, click

Use Last Zero

.

The status bar (lower left corner) of the window displays “Transmitting...” until calibration is finished. It then
displays “Zero Calibration Complete.”

Note

Note