Calibration

37

Figure 4-3. Two-Point Calibration wizard

4. Select the calibration type (see Figure 4-3), then click

Next

.

5. Enter the amount of test weight being used.
6. If applicable, select the Chains or Hooks are used during calibration check box and click

Next

.

7. Remove all weight from the scale and click

Calibrate Zero

. If you want to use the zero acquired from the last

press of the Zero button, click

Use Last Zero

.

The status bar (lower left corner) of the window displays Transmitting... until calibration is finished. It then
displays Zero Calibration Complete.

8. Click

Next

.

9. Add the test weight to the scale and click

Calibrate Span

.

10. If linear points were selected in Step 5, enter the applied weight in the text box and click

Measure

.

The Status will display

Success

when complete.

Figure 4-4. Two-Point Calibration wizard

11. Click

Next

.

12. If chains or hooks are being used, remove them (along with all other weight) from the scale and click

Re-Zero

Calibrate

. Click

Next

.

13. Calibration is complete. Click

Finish

to commit the calibration values and send them to iQUBE

2

.

Click

Cancel

if you want to abort. This will require a re-calibration at a later time.

4.2.2 Corner Match

This option allows you to calibrate and trim cells based on weight placed on individual load cells.

1. From the

Scales

section, select the scale to be calibrated.

2. From the

General

tab, select

Corner Match

from the

Calibration Type

pull-down menu.

3. Click the

Calibrate

button.

The Corner Match Calibration wizard appears.

Note