4.0 Calibration

Calibration types depend on how many load cells are connected to iQUBE

2

. See Table 4-1. If only one load cell is

connected, only 2-point calibration will be available.

The iQUBE

2

can be calibrated by using the Virtui

2

Configuration Utility

,

Revolution III, 920i, or serial commands.

No minimum test weight values are required; however, at least 10% of the scale capacity is recommended.

4.1

920i Calibration

Verify that the iQUBE

2

is connected to the 920i

with power applied. Use the 920i

setup switch to make sure the

indicator is in setup mode.

Do not move the

iQUBE

2

SW1 switch. It should remain in the OFF position (see Figure on page 6 and Figure 3-

1 on page 19). If

iQUBE

2

is not connected, menus may run very slowly. Configuration must be downloaded prior

to calibrating.

Do not mix calibration types (section, corner, 2-point). Span adjustments must be made in appropriate
calibration type.

Figure 4-1. 920i CALIB Submenu

Calibration Type

Description

2-POINT

Traditional span calibration. No trimming will be done.

CORNER

Used when weight will be placed over each load cell.

SECTION

Used when load cells are paired with each other and calibrated in those pairs.

THEORETICAL

Mathematically calibrates the scale using full scale capacity and load cell sensitivities.

Table 4-1. Calibration Types

Note

Important

DIAGNOS

GRADS

FORMAT

CALIB

SCALES

SERIAL

FEATURE

PFORMT

SETPTS

DIG I/O

VERS

SCALE x

WZERO

WSPAN

WLIN

WVAL

CALMATCH

CALSPAN

WLIN

POINT 1

POINT 2

weight_value

span value

count

WGT1

CAL1

Same as
Point 1

Same as Point 1

Diagnostics

Screen

For

Trimming

CALZERO

Last Zero or

Temp Zero

SECTION

2-POINT

POINT 3

POINT 4

POINT 5

POINT 1

POINT 2

CORNER

Same as

SECTION

Theoretical

Calibration

THEORET

POINT 4

POINT 5

POINT 3

Theoretical

Calibration

deadload

count value

weight_value

span value

count

Deadload

adjustment

REZERO

WVAL

THEORET

REZERO

Deadload

adjustment