Configuration

31

3.3

Revolution III Configuration

The Revolution III configuration utility can be used to set iQUBE

2

configuration parameters for use with any indi-

cator. When Revolution III configuration is complete, configuration data is downloaded to the iQUBE

2

.

Revolution III supports both uploading and downloading of iQUBE

2

configuration data. This capability allows con-

figuration data to be retrieved from one iQUBE

2

, edited, then downloaded to another. Revolution III provides

online help for each of its configuration displays.

To use Revolution III, do the following:

1. Install Revolution III on a PC running Windows

®

98 or later.

2. With both the iQUBE

2

and the PC powered off, connect the PC serial port to pins 1, 2, and 5 of connector J7

on the iQUBE

2

connector board. (See Table 2-2 on page 8).

3. Power up the PC and the iQUBE

2

. Use the setup switch to place the iQUBE

2

in setup mode.

4. Start the Revolution III program.

3.3.1 Revolution III Configuration

To configure the iQUBE

2

using Revolution:

1. With the PC and iQUBE

2

connected and

Revolution III running, select

New

from the

File

menu.

2. From the Select Indicator window, double-

click the iQUBE

2

icon, or select it and click

the

New Configuration File

radio button, then

click

OK

. The iQUBE

2

Information Display

is shown.

3. Select

Base Configuration

from the left pane

(see Figure 3-17) to expand it. The Base
Configuration
selections (

General, Load

Cells, Load Cell Assignment, Comm Ports, Digi-
tal I/O,

and

Diagnostics

) are displayed.

Primary/Secondary Configuration

1. From the Base Configuration list, on the left

pane, click

Primary Secondary

.

2. For each secondary device, enter the desired

eight-digit ID.

3. Set the address to a value other than 1 to

activate the device. Each secondary device
must have its own unique address.

To deactivate the secondary device, set
the address to 1 (use the address text boxes shown in Figure 3-17).

General Configuration

1. From the Base Configuration list, on the left pane, click

General

. Use the

Formatting

tab to modify the date

and time formats.

2. Use the

Sample Rate

tab to specify the sample rate (2.5HZ-500HZ).

Load Cell Configuration
The

Load Cells

section allows you to enter the capacity, sensitivity, name, serial number, zero value, and normaliza-

tion value for each load cell (up to 16).

1. From the left pane, click

Load Cells

.

2. From the listing of load cells, click on the one to be customized.
3. Using the text boxes, specify the capacity and sensitivity. Optionally, a load cell name and the load cell serial

number can be entered.

4. Zero and normalization values are set when the scale is calibrated, but can be customized by entering new

values in the text boxes.

5. Repeat Steps 1-4 for as many cells as are connected to the iQUBE

2

.

Figure 3-17. Rev III Base Configuration, Primary/Secondary

Communications

Note