4.3

Revolution III Calibration

For reference, Revolution III calibration examples are included on the CD, PN 77925. To calibrate the iQUBE

2

by

using Revolution III,

1. Ensure that you are connected to iQUBE

2

by clicking the

Connect

button on the toolbar or selecting

Commu-

nications » Connect

. The

Disconnect

button becomes active when you are connected.

Figure 4-7.

Revolution III

Toolbar, Connect Button

Figure 4-8.

Revolution III

Toolbar, Connected - Disconnect Button Enabled

2. From the

Tools

menu, select

Calibration Wizard...

Click

Calibrate

to begin calibration.

3. Select whether to begin a calibration, temporary zero calibration, or theoretical calibration. Click

Calibrate

to

begin a calibration.

4. Select whether you want to use a standard calibration, standard calibration with linear points, or span only

calibration and click

Next

.

5. In the text box, enter the amount of test weight being used.
6. If chains or hooks are being used, select the check box.
7. Remove all weight from the scale. If you selected Chains or Hooks are used during calibration in Step 6,

place them on the scale and click

Next

.

8. Click the

Calibrate Zero

button. If you want to use the zero acquired from the last press of the Zero button,

click

Use Last Zero

.

The status bar (lower left corner) of the window displays Transmitting... until calibration is finished. It then
displays Zero Calibration Complete.

9. Add the test weights specified in Step 5 to the scale.
10. Click the

Calibrate Span

button.

11. If you selected Chains or Hooks are used during calibration in Step 6, remove them from the scale and click

the

Click to Re-Zero

button.

12. If you selected Multi-Point Linearization in Step 4, use the Weight Value text boxes to enter test weight values

for each linear point. Then, click

GO

to calibrate each linear point.

Figure 4-9. Calibration Wizard, Linearization Points

13. Calibration is complete. Click

Finish

to commit the calibration values and send them to iQUBE

2

.

Click

Cancel

if you want to abort. This will require a re-calibration at a later time.