iQUBE

2

Installation Manual

6.1.2 Retrieving Diagnostic Data

Table 6-2 also lists the diagnostic query commands associated with each of the diagnostic codes. When a diagnos-
tic code is generated, the related query command can be used to retrieve additional information about the error.

Using the query command for a given error also clears the diagnostic flag for that error. Once cleared, that data
is no longer available.

6.2

Diagnostic Commands

Table 6-3 summarizes the diagnostic commands available for the iQUBE

2

, including query commands, parameter-

setting commands, and the values that can be assigned by the parameter-setting commands. The diagnostic com-
mands are described in detail in the following sections.

6.2.1 System Diagnostic Commands

System diagnostic commands consist of query commands only; there are no parameter-setting commands. System
diagnostics are always enabled.

For data formats that include the <S> status token, some system diagnostic errors also set the status to

I

(invalid).

Power Supply Error, <D> = P
The power supply diagnostic flag is set when any CPU board in the system has a power supply fault (input voltage
less than 6 VDC or greater than 12 VDC).

<D>

Error Type

Query Command

Parameter Setting

Commands

Parameter Values

Default

Value Range

P

Power supply

DIA.PS

S

Secondary
communication

DIA.MSCONNECT

E

Excitation

DIA.PSEXC

C

Cell connection

DIA.CELLCONNECT

V

Overloaded cell

DIA.OVERLOAD

SCs.DIA.OVERLOAD

100

–100 – 100
(percent of cell)

U

Underloaded cell

DIA.UNDERLOAD

SCs.DIA.UNDERLOAD

10

–100 – 100
(percent of cell)

R

Zero reference

DIA.ZREF

SCs.DIA.ZREF

OFF

OFF, ON

SCs.DIA.ZREF.RANGE

2

0 – 100 (percent of scale)

SCs.DIA.ZREF.THRESH

10

–100 – 100 (percent of
scale)

SCs.DIA.ZREF.TIME

10

1 – 60 (seconds)

D

Drift

DIA.DRIFT

SCs.DIA.DRIFT

OFF

OFF, ON

SCs.DIA.DRIFT.RANGE

2

0 – 100.0
(percent of scale)

SCs.DIA.DRIFT.THRESH

2

0 – 100.0
(percent of scale)

SCs.DIA.DRIFT.TIME

10

1 – 300 (seconds)

N

Peak noise

DIA.NOISE

SCs.DIA.NOISE

OFF

OFF, ON

DIA.NOISE.THRESH

10.00

0.01 – 1000 (

V)

SCs.DIA.NOISE.TIME

10

1 – 300 (seconds)

L

Unbalanced cell

DIA.UNBAL

SCs.DIA.UNBAL

OFF

OFF, ON

SCs.DIA.UNBAL.RANGE

5

5 – 75 (percent of scale)

SCs.DIA.UNBAL.THRESH

10

0 – 50 (percent of scale)

All

DIA.FLAGS

All

DIA.CLEAR

Table 6-3. Diagnostic Command Summary

Note

Note