Running the Stability Test from the Web Interface

A Stability Test column on the IT test results display (see picture above) shows PASS or FAIL for each load sensor.
We obtained the data by running the IT test, as described above.

Stability test is in the Diagnostics menu. Select the Stability Test hyperlink and wait a few seconds.

The results will not be specific to each load sensor, as in the IT test. They will display an average reading and
PASS/FAIL for the system as a whole.If equipped with an IT summing junction box there will be a selection to run
the Weight and voltage test. That test will display the individual load cell and loop back test results.

Running the Stability Test from the Optional Front Panel

1) From the Diagnostic menu, select Stability Test with use the down arrow and press Enter. The HI 6600 runs

the test and shows the results.

2) To see the AD/Raw Count and the A/D Filtered select them from the menu.

3) Press Exit to return to the Test Menu.

Stability:

Mean
Variation
Results
mV/V
RTZ

Gross = 0.9lb

RAW
32918
0.17
OK
0.0184
Pass

Waversaver
32993
0.00