Hardy HI 6600 Series Modular Sensor System User Manual

Page 60

Advertising
background image

HI 6600 Series Modular Sensor System User Guide

Page | 60

7) From the DISPLAY drop down, select the Long Inverse selection. This will allow us to write an integer value

into the non-float registers.

8) From the FUNCTION drop down list select Read/Write registers, or click button 23 to open the Write

multiple registers display.

9) Double click on the first register and the Enter Value box appears. Enter the new value you wish to write to

this register. Our example shows writing a value of “2”, which is the Tare command number.