C2 and eCAL Calibration Process

Navigate to the C2and eCAL menu from the Web brower or on the front panel display

1) Check to make sure the number of load cells displayed in the Num Devices menu corresponds to

the actual number of C2 devices installed. If the number varies, check to ensure that each
load cell/point cable connection is securely fastened and that each load cell/point cable is not
broken. Only Hardy Process Solutions load sensors are C2 (or eCal) calibration capable.

2) In the “Ref Weight” text field, enter the reference weight for your application. An empty scale

will use a 0.00 reference weight setting.

3) Enter the Gravity Correction Factor, see the table above, then enter the Correction Factor

number you selected from the table in the Gravity Correction field that best corresponds to
your location.

4) Click the “Do C2 Calibration” button.

Wait a few seconds and the results will appear.

If the Calibration was successful a Cal completed OK appears and gross weight will read your reference
value.

If the Calibration failed, check to make sure C2 load cells are being use and properly connected.
If the system is using one or more non-C2 load cells, calibrate using a traditional calibration (Hard Cal)
process.

See the Troubleshooting section of this manual for further information.


5) Click “back” in the Web browser or use the left arrow key on the front panel display to return to the main

Calibration menus.