Count operations – Hardy HI 6600 Series Modular Sensor System User Manual

Page 46

Advertising
background image

HI 6600 Series Modular Sensor System User Guide

Page | 46

Count Operations

The count mode allows the unit to be calibrated and used as a piece counter. This mode allows the user to "count" the
number of pieces placed on the scale.

Piece Count can be displayed on the main screen by first enabling the count mode in the menus, then pressing the MODE
button to cycle through showing: GROSS > NET > COUNT >

Determining Piece Count:

If the weight of an individual piece is known:

1)

First select Count Enabled

2)

Select Unit Weigh by pressing ENTER and input the single unit weight.

If the weight of an individual piece (or pieces) is unknown:

1) Place a sample or samples onto the scale and enter the number of pieces being used for the "sample size".

2) Press or click on the “weigh sample”

The scale will divide the total weight of the “weighed sample” by the “sample size” to determine the weight of a single
unit.

NOTE All the samples must be the same weight, type and size.

Be sure the scale has been Zero’d before determining piece count.

Hint:

When the weight of individual pieces varies, it is better to take the Weight Sample of a much larger
number of pieces. This allows the individual pieces to have greater variability but ensures an
accurate count.