HI 6600 Series Modular Sensor System User Guide

Page | 117

0x64 (100 decimal): CAL LOW CMD. Write a 0x64 hex to the command register to perform the
low step of a traditional calibration. The status register will read 0 if this command succeeds:

o

Calibration_Fail 1

o

Calibration_Fail_Motion 3

o

Calibration_Fail_Adc_Error 4

0x65 (101decimal): CAL HIGH CMD. Write a 0x65 hex to the command register to perform the
high step of a traditional calibration.

o

Calibration_Fail 1

o

Calibration_Fail_Motion 3

o

Calibration_Fail_Adc_Error 4

o

HardcalFailCounts 8: not enough counts between hard cal hi and hard cal lo

0x66 (102 decimal): C2 CAL CMD. Write a 0x66 hex to the command register to perform a C2
calibration.

o

Calibration_Fail 1

o

Calibration_Fail_Motion 3

o

Calibration_Fail_Adc_Error 4

o

Calibration_Fail_Noc2 5

o

Calibration_Fail_C2capeq 6

o

Calibration_Fail_C2clones 7

0x1000 (4096 decimal): WRITE INTEGER CMD. Set the value of an integer parameter.
Write 0x1000 in the command register 0, the parameter ID number in registers 2, 3 and the
desired value in registers 4, 5.

o

No Error Codes

0x1001 (4097 decimal): WRITE FLOAT CMD. Set the value of a floating point parameter.
Write 0x1001 in the command register 0, the parameter ID number in registers 2, 3 and the
desired float value in registers 4, 5.

• Failure 1
• TooHigh -1 : value is above legal limit
• TooLow -2 : value is below legal limit
• ErrorNotFound 0x8000 : parameter id not found