Setup Parameter Menus

The Setup Menus consists of the following parameters:

 Capacity
 Decimal Point
 Grads
 Instrument ID
 Motion Tolerance
 Operator ID
 Unit

Capacity Parameter

Scale Capacity is the scale's nominal operating capacity (the total weight capacity of the scale system). If this value is
exceeded by six graduations, dashes appear on the front display. Communications to and from optional devices are not
affected.

If a Capacity value is entered which conflicts with either the decimal point value or the graduation value, the decimal
point and/or graduation values are automatically modified to match the programmed capacity. It is therefore
recommended that the Capacity parameter is entered first before setting the Decimal Point and Grad parameters.

RANGE: .000001 - 999999 (Default 999999).

On the Web page Or Using the Front Panel Display: enter the capacity in the text box provided or enter the capacity
using the arrow keys.

Decimal Point Parameter

Use the Decimal Point Parameter to set the resolution you want for the WPM.Here you set the location of the decimal point for
the weight resolution. The higher the number, the farther to the left the decimal point moves and the higher the resolution of the
scale. Note that setting more decimal points does affect the overall accuracy of the instrument.

RANGE: 0-5 (default 2) Not to exceed 1 part in 100,000

On the Web page Or Using the Front Panel Display: from the Decimal Point pull-down list, select the decimal position for
this instrument or select from the list of decimal point values supported using the arrow keys.

Graduation Size Parameter

The Graduation Size is the Minimum increment displayed by the instrument. The Base Graduation Number can be
calculated by dividing the Total Load Cell Capacity by 10,000.

For example, with two decimal points selected, a graduation size of 10 will display increments of .10 units and the
graduation size .50 will display increments of.50 units. For a scale with 10,000 capacity, graduation size = 1
RANGE:1, 2, 5,10, 20, 50, 10, 200, 500, 1000 (default 1)

On the Web page Or Using the Front Panel Display: from the Decimal Point pull-down list, select the decimal position
for this instrument or select from the list of graduation values supported using the arrow keys.