Common Error Messages

 A/D Convert Error! - Load Cells input out of range.

 Motion Error! - Check Motion Tolerance Settings and Retry

 Too Lo Error! - Verify that the load cell signal level is 0-25 mV. Verify that there is enough weight on

the scale.

 Too Hi Error! - Verify that the load cell signal level is 0-25mV. Verify that there is too much weight on

the scale.

 No C2 Sensor! - Instrument did not detect a C2 Load Sensor

 CAL Failed! - Too few counts between Zero and Span.

 C2 Caps Unequal! - Different load cell capacities (For example 50 lbs capacity load cell and 100 lbs capacity load cell

on one system. Make the load cells equal by removing the odd load cell and replacing it with a load cell that is equal to
the other’s capacity.

 HI/LO Too Close! - Zero and Span are not more than 1,000 counts from each other or there is no change or negative

change. Reset either so the counts are more than 1,000 counts of each other.

 Not Allowed! - Value entered is outside the range allowed. Try another value.

 Need Cal with ITJBOX - IT summing card is not installed. Install an IT summing card then do a Calibration with the

card installed to access the IT information.

Diagnostics

Diagnostics is covered extensively in the Troubleshooting Section (Chapter 8). The available parameters on the Web

page are:

 Parameters (which shows all parameters on the instrument)
 Integrated Technician (when connected to an IT Junction Box)
 C2
 Stability Test

The available parameters on the device are:

 Stability Test
 Weight and Voltage Test

Trouble Shooting Using Integrated Technician (IT

®

)

The Optional Front Panel display for the HI 6600 instrument’s Stability test is shown below. To navigate this screen, select a test
on the right using the right and left arrows with exit using the left arrow when you are at the first out of the five positions.

The test displays are live readings and only available if you have an IT summing junction card, maximum of one J-box and 4