HI 6600 Series Modular Sensor System User Guide

Page | 83

Modifying the Calibration Parameters

The calibration security locks out users from seeing any calibration parameters until you enter the correct calibration
password.

To modify the calibration parameters use the UP or DOWN button to select the calibration menu and press the ENTER
button. The password dialog box will appear; if the correct 4 digit alphanumeric character sequence is entered the
calibration lock will be disabled until you exit the calibration menu; at which time the calibration lock will be
automatically reset to the original setting.

If an incorrect password is entered the menu will return to the configuration menu.

Modifying the Read Only Parameters

The security and calibration parameters have their own password protection. The parameters discussed in this section are
those listed under communication, diagnostics, display, filter, language, operation, and setup.

With the exception of the security and calibration parameters, all parameters can be read by the operator once the operator
is navigating inside the configuration menu; however the ability to change these parameters can be password protected.

If the read only password is enabled and a parameter is selected, the password dialog box will appear. If the 4
alphanumeric character sequence is entered the read only lock will be disabled until you exit the configuration menu; at
which time the configuration lock will be automatically reset to the original setting.

If an incorrect password is entered, the menu will return to the configuration menu.