Table 8-1, D table 8-2, Ed table 8-1 f – Emerson OXYMITTER 4000 User Manual

Page 108: Table 8-2, Diagnostic/unit alarm fault definitions – loi -4, Ted in table 8-1, In table 8-2, Oxymitter 4000

Advertising
background image

Instruction Manual

IB-106-340 Rev. 3.0
December 2003

8-4 Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management

Oxymitter 4000

Table 8-1. Diagnostic/Unit Alarm Fault Definitions – Membrane Keypad Only

LED

Flashes

Status

4-20 mA Line

Fault

Self-

Clearing

HEATER T/C

1

OPEN

Dependent on position 3 of SW2*

1

NO

2

SHORTED

Dependent on position 3 of SW2*

2

NO

3

REVERSED

Dependent on position 3 of SW2*

3

NO

4

A/D COMM ERROR

Dependent on position 3 of SW2*

4

NO

HEATER

1

OPEN

Dependent on position 3 of SW2*

5

NO

2

HIGH HIGH TEMP

Dependent on position 3 of SW2*

6

NO

3

HIGH CASE TEMP

Dependent on position 3 of SW2*

7

YES

4

LOW TEMP

Dependent on position 3 of SW2*

8

YES

5

HIGH TEMP

Dependent on position 3 of SW2*

9

YES

O2 CELL

1

HIGH mV

Dependent on position 3 of SW2*

10

YES

3

BAD

Track O

2

11

YES

4

EEPROM CORRUPT

Dependent on position 3 of SW2*

12

NO

CALIBRATION

1

INVALID SLOPE

Track O

2

13

YES

2

INVALID CONSTANT

Track O

2

14

YES

3

LAST CALIBRATION
FAILED

Track O

2

15

YES

**

CALIBRATION RECOM-
MENDED

Track O

2

YES

*Critical alarm conditions will render the O

2

measurement as unusable, and any of these events will cause the

4-20 mA signal to go to a user-selectable limit of 3.5 mA or 21.6 mA (position 3 of SW2). Factory default value
is 3.5 mA. Alarms which are not “self-clearing” will require recycling of power to the electronics.

**The CALIBRATION RECOMMENDED alarm flashes the Calibration Recommended alarm LED on the opera-

tor’s keypad.

Table 8-2. Diagnostic/Unit Alarm Fault Definitions – LOI

Message

Status

Fault

Number

Self-

Clearing

O2 T/C OPEN

HEATER T/C OPEN

1

NO

O2 T/C SHORTED

HEATER T/C SHORTED

2

NO

O2 T/C REVERSED

HEATER T/C POLARITY REVERSED

3

NO

ADC ERROR

A/D COMM ERROR

4

NO

O2 HEATER OPEN

O2 HEATER OPEN

5

NO

VERY HI O2 TEMP

VERY HIGH PROCESS TEMPERATURE

6

NO

BOARD TEMP HI

ELECTRONICS OVERHEATED

7

YES

O2 TEMP LOW

LOW PROCESS TEMPERATURE

8

YES

O2 TEMP HI

HIGH PROCESS TEMPERATURE

9

YES

O2 CELL OPEN

O2 CELL OPEN

10

YES

O2 CELL BAD

O2 CELL FAILED

11,13, 14

YES

EEPROM CORRUPT

EEPROM FAILED

12

NO

CALIB FAILED

LAST CALIBRATION FAILED

15

YES

LINE FREQ ERROR

INCORRECT INPUT LINE FREQUENCY
DETECTED ON POWER UP

NO