Table 8-3, Sps 4000 fault finding -22, To table 8-3 – Emerson OXYMITTER 4000 User Manual

Page 126: Oxymitter 4000

Advertising
background image

Instruction Manual

IB-106-340 Rev. 3.0
December 2003

8-22 Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management

Oxymitter 4000

Table 8-3. SPS 4000 Fault Finding

Symptom

Check

Fault

Remedy

No calibration
gas flow

Wiring

Improper wire connections,
loose connections, or
damaged wiring

Properly connect wiring or secure
loose wiring connections; replace
damaged wiring if necessary.

Logic I/O

Oxymitter 4000 logic I/O
not set for calibration
handshaking with SPS
4000

Set logic I/O to mode 8 via
HART/AMS.

Calibration gas lines be-
tween cylinders and mani-
fold

Clogged calibration gas
line

Replace clogged calibration gas line.

Calibration gas flowmeter
knob

Flowmeter knob not turned
counterclockwise to allow
flow

Turn calibration gas flowmeter knob
counterclockwise to allow calibration
gas to flow.

Calibration gas line be-
tween manifold and cali-
bration gas flowmeter

Clogged calibration gas
line

Replace clogged calibration gas line.

Fuse on power supply
board

Blown fuse

Replace fuse per paragraph 9-5a.

Interface board operation

Interface board not send-
ing signals

Replace interface board per para-
graph 9-5b.

Check valve

Clogged check valve

Replace check valve per paragraph
9-5e.

Calibration gas line be-
tween calibration gas
flowmeter and check valve

Clogged calibration gas
line

Replace calibration gas line.

Calibration gas flowmeter

Clogged flowmeter

Replace flowmeter per paragraph
9-5h.

Power supply output

Power supply failure

Replace power supply board per
paragraph 9-5b.

Solenoid

Solenoid failure

Replace solenoid per paragraph
9-5c.

Pressure switch

Pressure switch failure

Replace pressure switch per para-
graph 9-5d.

Advertising