1-14 Description and Specifications

Rosemount Analytical Inc. A Division of Emerson Process Management

Oxymitter 4000

1-7 SPECIFICATIONS

Oxymitter 4000

O

2

Range:

Standard ..................................................

0 to 10% O

2

,

0 to 25% O

2

,

0 to 40% O

2

(via HART)

Accuracy ..................................................

±0.75% of reading or 0.05% O

2

, whichever is greater

System Response to Calibration Gas ......

Initial – less than 3 seconds, T90 – less than 8 seconds

Temperature Limits:

Process ....................................................

32° to 1300°F (0° to 704°C) up to 2400°F (1300°C) with
optional accessories

Electronics ...............................................

-40° to 185°F (-40° to 85°C) [Operating temperature of
electronics inside of instrument housing, as measured by a
HART communicator or Rosemount Asset Management
Solutions software.]

Local Operator Interface ..........................

-40° to 185°F (-40° to 85°C) [Above 158°F (70°C), the in-
frared keypad will cease to function, but the Oxymitter
4000 will continue to operate properly.]

Probe Lengths ...............................................

18 in. (457 mm)

12 ft (3.66 m)

3 ft (0.91 m)

15 ft (4.57 m)

6 ft (1.83 m)

18 ft (5.49 m)

9 ft (2.74 m)

Mounting and Mounting Position ...................

Vertical or horizontal
Spool pieces, P/N 3D39761G02, are available to offset
transmitter housing from hot ductwork.

Materials:

Probe .......................................................

Wetted or welded parts - 316L stainless steel (SS)
Non-wetted parts - 304 SS, low-copper aluminum

Electronics Enclosure ..............................

Low-copper aluminum

Calibration .....................................................

Manual, semi-automatic, or automatic

Calibration Gas Mixtures Recommended .....

0.4% O

2

, Balance N

2

8% O

2

, Balance N

2

Calibration Gas Flow .....................................

5 scfh (2.5 l/m)

Reference Air ................................................

2 scfh (1 l/m), clean, dry, instrument-quality air
(20.95% O

2

), regulated to 5 psi (34 kPa)

Electronics .....................................................

NEMA 4X, IP66 with fitting and pipe on reference exhaust
port to clear dry atmosphere

Electrical Noise ..............................................

Meets EN 55022 Generic Emissions Std.,
Includes EN 61000-4-2 for Electrostatic Discharge
Includes EN 61000-4-3 for RFI
Includes EN 61000-4-4 for Fast Transients
Includes EN 61000-4-6 for RFI

Line Voltage ..................................................

90-250 VAC, 48/62 Hz. No configuration necessary. 3/4 in.
- 14 NPT conduit port

Signals:

Analog Output/HART ...............................

4-20 mA isolated from power supply, 950 ohms max. load

Logic I/O ..................................................

Two-terminal logic contact configurable as either an alarm
output or as a bi-directional calibration handshake signal to
IMPS 4000 or SPS 4000