Instruction Manual

IB-106-340 Rev. 3.0

December 2003

Rosemount Analytical Inc. A Division of Emerson Process Management

Configuration with LOI 4-3

Oxymitter 4000

f. Once the cell is up to operating temperature,

the O

2

percentage can be read:

1. To access TP5 and TP6 next to the

LOI (Figure 4-2), power down the
Oxymitter 4000 and remove the LOI
module. Attach alligator leads from a
multimeter across TP5 and TP6. Install
the LOI module and power up the
Oxymitter 4000. Allow time for the cell
to reach operating temperature. The
calibration and process gases can now

be monitored. When a calibration has
been initiated, the value at TP5 and
TP6 is the % O

2

seen by the cell. Oxy-

gen levels, as seen on the multimeter,
are:

8.0% O2 = 8.0 VDC
0.4% O2 = 0.4 VDC

2. HART/AMS.

3. Model 751. The loop-driven LCD

display.

SW2

SW1

1

2
3
4

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

4-20 mA

IS INTERNALLY

POWERED

(DEFAULT)

4-20 mA REQUIRES
EXTERNAL POWER

HART

0 TO 10% O

2

3.5 mA

220 V

115 V

0 TO 25% O

2

LOCAL

21.6 mA

DEFAULT

POSITION

(EX-FACTORY)

3.5 mA/21.6 mA:

0 TO 25% O :

2

0 TO 10% O /

2

LOCAL:

HART: O RANGE SET BY HART/AMS

(FROM 0 TO 40% O )
O RANGE SET BY POS 2

O RANGE
WHEN ALARM EXISTS, OR
ON POWER UP, CURRENT
OUTPUT GOES TO THIS VALUE

2

2

2

2

37260012

ON

OFF

NOTE:

THE 115 V OPTION
AT SWITCH SW2
POSITION 4 IS ACTIVE
ONLY WHEN THE
HEATER VOLTAGE
OPTION IS SET TO
MANUAL IN THE
SOFTWARE.

Figure 4-2. Defaults – Oxymitter 4000 with LOI