Figure 8-5. fault 3, reversed thermocouple -7, Oxymitter 4000, Alarms o2 t/c reversed – Emerson OXYMITTER 4000 User Manual

Page 111: Instruction manual

Advertising
background image

Instruction Manual

IB-106-340 Rev. 3.0

December 2003

Rosemount Analytical Inc. A Division of Emerson Process Management

Troubleshooting 8-7

Oxymitter 4000

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

37260021

Alarms

O2 T/C Reversed

LOI

KEYPAD

Figure 8-5. Fault 3, Reversed Thermocouple

c. Fault 3, Reversed Thermocouple Wiring or

Faulty PC Board

Figure 8-5 shows the electronic assembly
for an Oxymitter 4000 with a membrane
keypad (upper view) and an Oxymitter 4000
with an LOI (lower view). The upper view
also shows J1 and test points TP1 through
TP6, located on the microprocessor board,
below the membrane keypad or the LOI
module.

Membrane Keypad. When Fault 3 is de-
tected, the HEATER T/C LED flashes three
times, pauses for three seconds, and re-
peats.

1. Using a multimeter, measure TP3+ to

TP4-.

2. If the reading is negative, the thermo-

couple wiring is reversed.

3. Check red and yellow wires in the J1

connector for the proper placement.

4. If the wiring is correct, the fault is in the

PC board. See paragraph 9-4d, Elec-
tronic Assembly Replacement.

LOI. When Fault 3 is detected, the LOI dis-
plays the “O2 T/C Reversed” message.

1. Remove power. Unscrew and remove

the LOI module from the electronic as-
sembly.

2. Reconnect power to the Oxymitter

4000.

3. Perform the diagnostic steps 1 through

4 shown for the membrane keypad.