Figure 8-12. fault 10, high cell mv -14, Oxymitter 4000, Alarms o2 cell open – Emerson OXYMITTER 4000 User Manual

Page 118: Instruction manual

Advertising
background image

Instruction Manual

IB-106-340 Rev. 3.0
December 2003

8-14 Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management

Oxymitter 4000

Alarms

O2 Cell Open

LOI

KEYPAD

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

37260028

Figure 8-12. Fault 10, High Cell mV

j. Fault 10, High Cell mV

Figure 8-12 shows the electronic assembly
for an Oxymitter 4000 with a membrane
keypad (upper view) and an Oxymitter 4000
with an LOI (lower view). The upper view
also shows J1 and test points TP1 through
TP6, located on the microprocessor board,
below the membrane keypad or the LOI
module.

Membrane Keypad. When Fault 10 is de-
tected, the O2 CELL flashes once, pauses
for three seconds, and repeats.

1. Using a multimeter, measure across

TP1+ to TP2-.

2. If you measure 1.2 VDC, the cell wires,

either orange or green, have become
detached from the input.

3. One possible cause is connector J1.

The orange or green wire has come
loose from the crimped connection.

4. The platinum pad could also be at fault.

The pad could have broken free from
the back of the cell.

5. Replace heater strut per paragraph

9-4h, Heater Strut Replacement. If
necessary, replace the cell flange
assembly per paragraph 9-4i, Cell
Replacement.

LOI. When Fault 10 is detected, the LOI
displays the “O2 Cell Open message.

1. Remove power. Unscrew and remove

the LOI module from the electronic as-
sembly.

2. Reconnect power to the Oxymitter

4000.

3. Perform the diagnostic steps 1 through

5 shown for the membrane keypad.